Literature DB >> 7377924

How to increase the stability of external fixation units. Mechanical tests and theoretical studies.

E Egkher, H Martinek, B Wielke.   

Abstract

From general physical laws simple formulae have been deduced from which the rigidity of an external fixation unit mounted in different ways becomes mathematically calculable. Furthermore, the stability of various mounting systems under bending forces was investigated experimentally. Particular attention was paid to details of operating technique which are of special importance to the stability of the whole construction. The calculated data are consistent with the mechanical test results. The best arrangement of the Steinmann pins and Schanz screws is discussed, and how to mount the frame optimally, depending on the specific type and location of the fracture. Special instructions concerning the mounting of the external fixation unit are given which, depending on the specific circumstances, guarantee optimum stability of the unit.

Entities:  

Mesh:

Year:  1980        PMID: 7377924     DOI: 10.1007/bf01246140

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0344-8444


  3 in total

1.  [Stability and strain of osteosynthesis of thigh and tibia in early mobilization (author's transl)].

Authors:  K Diehl
Journal:  Unfallheilkunde       Date:  1976-03

2.  [Biomechanics of the external fixation clamps].

Authors:  R Labitzke; G Henze
Journal:  Unfallheilkunde       Date:  1978-08

3.  External fixation. Classification and indications.

Authors:  G Hierholzer; R Kleining; G Hörster; P Zemenides
Journal:  Arch Orthop Trauma Surg       Date:  1978-08-30
  3 in total
  2 in total

1.  Stability of the original Hoffman and AO tubular external fixation devices.

Authors:  T K Moroz; J B Finlay; C H Rorabeck; R B Bourne
Journal:  Med Biol Eng Comput       Date:  1988-05       Impact factor: 2.602

2.  Hughes external fixator in treatment of tibial fractures.

Authors:  C M Court-Brown; S P Hughes
Journal:  J R Soc Med       Date:  1985-10       Impact factor: 5.344

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.