A comparison of fluorescent and nonfluorescent light sources for phototherapy.

J B Warshaw, J Gagliardi, A Patel.   

Abstract

We have compared fluorescent and nonfluorescent light sources for phototherapy for newborn infants with hyperbilirubinemia. Phototherapy was provided by a tungsten halogen lamp and conventional fluorescent light with identical radiant flux of 6 microW/sq cm. For 22 infants treated with the nonfluorescent lamp the mean duration of phototherapy was 33.77 hours and the mean reduction of bilirubin was 3.84 mg/100 ml/day. This did not differ significantly from infants treated with conventional fluorescent lights. The nonfluorescent light can be utilized for infants in incubators or on radiant warmers. These results provide additional support for the relationship between radiant flux as a practical measure of phototherapy dose and the clinical response of a reduction in serum bilirubin.

Entities:  

Mesh:

Year:  1980        PMID: 7367087

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  3 in total

1.  Hyperbilirubinemia in term newborn infants. The Canadian Paediatric Society.

Authors: 
Journal:  Can Fam Physician       Date:  1999-11       Impact factor: 3.275

2.  Approach to the management of hyperbilirubinemia in term newborn infants.

Authors: 
Journal:  Paediatr Child Health       Date:  1999-03       Impact factor: 2.253

Review 3.  Use of phototherapy for neonatal hyperbilirubinemia. Fetus and Newborn Committee, Canadian Paediatric Society.

Authors: 
Journal:  CMAJ       Date:  1986-06-01       Impact factor: 8.262