Analytic expressions were derived for the response time, flow dependency, and signal-to-background ratio of blood gas catheters. These expressions were utilized to optimize the design of a new catheter. The catheter was tested in vitro and in vivo, and shown to be fast responding and practically flow independent, with an acceptable signal-to-background ratio.Entities:
Mesh:
Substances:
Year: 1980 PMID: 7364621 DOI: 10.1152/jappl.1980.48.2.376