| Literature DB >> 7363148 |
Abstract
The flow of oxygen to an anaesthetic gas machine suddenly and completely stopped during the course of a general anaesthetic. The cause of the failure was traced to a Schrader oxygen coupler check unit. During servicing, one component of the check unit had inadvertently been replaced by a similar but inappropriate component. Interruption of oxygen flow resulted from a loss of contact between two components of the check unit. This incident raises questions about the basic design and safety of Schrader coupler check units.Entities:
Mesh:
Substances:
Year: 1980 PMID: 7363148 DOI: 10.1007/bf03007784
Source DB: PubMed Journal: Can Anaesth Soc J ISSN: 0008-2856