Abstract
Two disposable condenser humidifiers were evaluated in nine ventilated intensive care patients. The Portex "Humid Vent" delivered end-inspired absolute humidities of 22--26.3 g/m3 at end-inspired temperatures of 27--28.3 degrees C. Corresponding humidities and temperatures with the Servo "Humidifier 150" were higher and were constant, at 27.7--29 g/m3 and 29.3--29.7 degrees C respectively. These disposable devices can be used for humidification in intensive care, but only for patients breathing room air, or on a short term basis.Entities:
Mesh:
Year: 1981 PMID: 7316137 DOI: 10.1177/0310057X8100900404
Source DB: PubMed Journal: Anaesth Intensive Care ISSN: 0310-057X Impact factor: 1.669