[Comparative investigations on the stability of fracture fixations with different fixateurs externes (author's transl)].

L Claes, C Burri, H Gerngross.   

Abstract

Short oblique fractures of the tibiae were stabilized with two-dimensional (frame spanner), three-dimensional and various V-shaped fixateurs externes, applied at the ventral-medial side. The stability of all these different types of fracture fixations was tested with and without an additional lag screw. Under a torsional load the stability of the area of osteotomy was measured. The three-dimensional fixateur externe and the V-shaped fixateur externe yielded the best results. The use of an additional lag screw produced a greater and significant (p less than 0,01) improvement.

Entities:  

Mesh:

Year:  1981        PMID: 7303302     DOI: 10.1007/bf02589695

Source DB:  PubMed          Journal:  Unfallchirurgie        ISSN: 0340-2649


  2 in total

1.  [Fixateurs externes for fractures combined with damage to the soft tissue (author's transl)].

Authors:  G Hierholzer
Journal:  Langenbecks Arch Chir       Date:  1975-11

2.  [Biomechanics of the external fixation clamps].

Authors:  R Labitzke; G Henze
Journal:  Unfallheilkunde       Date:  1978-08
  2 in total
  4 in total

1.  [Intramedullary nailing as secondary intervention following prior use of external fixation].

Authors:  J Ahlers; G Ritter; H Weigand
Journal:  Unfallchirurgie       Date:  1983-04

2.  [Modified Schanz screws for the external fixator].

Authors:  D Hofmann; H Burger; K E Rehm; A Steiner; K Henneking
Journal:  Unfallchirurgie       Date:  1985-06

3.  [Treatment guidelines and results in severe open tibial fractures].

Authors:  W Scharf; E Orthner; M Wagner
Journal:  Unfallchirurgie       Date:  1984-08

4.  [Histologic finding of fracture healing using external fixation and its clinical significance].

Authors:  K M Stürmer
Journal:  Unfallchirurgie       Date:  1984-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.