Literature DB >> 7296841

Determination of hemoglobin derivatives with IL 282 CO-oximeter as compared with a manual spectrophotometric five-wavelength method.

A Zwart, A Buursma, B Oeseburg, W G Zijlstra.   

Abstract

Hemoglobin derivatives as determined with the IL 282 CO-Oximeter correlated well with results by a manual five-wavelength method, which in turn had been checked against established methods for one or two derivatives. Measurement of total hemoglobin yielded almost identical results with both methods. As for oxyhemoglobin, carboxyhemoglobin, and hemiglobin, agreement between the two methods was fair. Although the IL 282 CO-Oximeter has not been constructed for the determination of sulfhemoglobin, it appeared that the instrument can still give a strong indication as to the presence of this hemoglobin derivative. Results from the IL 282 for fetal human blood should be used with caution, especially because of the possibility of falsely high HbCO readings.

Entities:  

Mesh:

Substances:

Year:  1981        PMID: 7296841

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  4 in total

Review 1.  Pulse oximetry.

Authors:  J F Kelleher
Journal:  J Clin Monit       Date:  1989-01

2.  Bronchodilator action of inhaled nitric oxide in guinea pigs.

Authors:  P M Dupuy; S A Shore; J M Drazen; C Frostell; W A Hill; W M Zapol
Journal:  J Clin Invest       Date:  1992-08       Impact factor: 14.808

3.  Microwaving blood as a non-destructive technique for haemoglobin measurements on microlitre samples.

Authors:  Toby H Basey-Fisher; Nadia Guerra; Chiara Triulzi; Andrew Gregory; Stephen M Hanham; Molly M Stevens; Stefan A Maier; Norbert Klein
Journal:  Adv Healthc Mater       Date:  2013-09-04       Impact factor: 9.933

4.  Accuracy and utility of a new reflectance pulse oximeter for fetal monitoring during labor.

Authors:  A Izumi; H Minakami; I Sato
Journal:  J Clin Monit       Date:  1997-03