Literature DB >> 7284892

A hazard of scavenger port design.

R M Flowerdew.   

Abstract

This case report describes an inadvertent misconnection of the breathing and scavenging hoses on the anaesthesia machine which resulted in complete expiratory obstruction. The features which facilitate such a misconnection are described and modifications to reduce the likelihood of this event recurring are suggested.

Mesh:

Substances:

Year:  1981        PMID: 7284892     DOI: 10.1007/BF03010362

Source DB:  PubMed          Journal:  Can Anaesth Soc J        ISSN: 0008-2856