Double lumen needle in unipuncture dialysis type double headpump.

R Vanholder, M De Paepe, N A Hoenich, S Ringoir.   

Abstract

Single needle dialysis is used as a method of vascular access in many European dialysis centres. However, recirculation has been incriminated as a disadvantage of this procedure. According to in vitro studies, recirculation should be limited to a minimum when using double lumen needles. The present study makes an in vitro comparison between single and double lumen needles in unipuncture (single needle) dialysis of the type double headpump (Bellco, BL 760) and studies the role of recirculation in dialysis performance. Although recirculation is significantly reduced with the double lumen needle, compared to the single lumen needle, the clearance of small molecules is not significantly different for both needle types.

Mesh:

Year:  1981        PMID: 7275338

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595