Evaluation of urine culture screening by light-scatter photometry.

D C Hale, L D Thrupp, J M Matsen.   

Abstract

Urine screening for bacteriuria by light-scatter photometry (Autobac) was evaluated for accuracy and compared with a colony count by the calibrated loop method. Incubation time, inoculum size, precision, and interference of particulate matter were evaluated in an effort to standardize the screening procedure. Results showed that urines could be accurately screened for Enterobacteriaceae by inoculating a single Autobac cuvette chamber with 0.1 or 0.2 ml of urine and determining the voltage change after four hours. A change of greater than or equal to 0.2 units indicates significant bacteriuria. Decreased accuracy was noted for urines having greater than 10(5) cfu/ml of Pseudomonas species or gram-positive cocci, possibly because these organisms grow more slowly.

Mesh:

Year:  1981        PMID: 7270499     DOI: 10.1093/ajcp/76.2.208

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  1 in total

1.  Evaluation of a two-minute test for urine screening.

Authors:  M T Pezzlo; M A Wetkowski; E M Peterson; L M de la Maza
Journal:  J Clin Microbiol       Date:  1983-09       Impact factor: 5.948