| Literature DB >> 7270499 |
D C Hale, L D Thrupp, J M Matsen.
Abstract
Urine screening for bacteriuria by light-scatter photometry (Autobac) was evaluated for accuracy and compared with a colony count by the calibrated loop method. Incubation time, inoculum size, precision, and interference of particulate matter were evaluated in an effort to standardize the screening procedure. Results showed that urines could be accurately screened for Enterobacteriaceae by inoculating a single Autobac cuvette chamber with 0.1 or 0.2 ml of urine and determining the voltage change after four hours. A change of greater than or equal to 0.2 units indicates significant bacteriuria. Decreased accuracy was noted for urines having greater than 10(5) cfu/ml of Pseudomonas species or gram-positive cocci, possibly because these organisms grow more slowly.Mesh:
Year: 1981 PMID: 7270499 DOI: 10.1093/ajcp/76.2.208
Source DB: PubMed Journal: Am J Clin Pathol ISSN: 0002-9173 Impact factor: 2.493