Literature DB >> 7258155

Clinical evaluation of the Abbott MS-2 antimicrobial susceptibility testing system.

D L Sewell, M T Makler.   

Abstract

Antimicrobial susceptibilities were determined for 845 clinical isolates by the standardized Bauer-Kirby disk diffusion method and the automated Abbott Ms-2 system. There was an overall correlation of 90% between the two methods. The Abbott MS-2 susceptibility instrument offers a rapid and accurate method for the antimicrobial susceptibility testing of bacteria.

Entities:  

Mesh:

Substances:

Year:  1981        PMID: 7258155     DOI: 10.1093/ajcp/76.1.82

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  3 in total

1.  Rapid antimicrobial susceptibility testing of Gram-negative clinical isolates with the AutoMicrobic system.

Authors:  B A Backes; S J Cavalieri; J T Rudrik; E M Britt
Journal:  J Clin Microbiol       Date:  1984-06       Impact factor: 5.948

2.  Direct identification and susceptibility testing of gram-negative bacilli from BACTEC bottles by use of the MS-2 system with updated bacterial identification software.

Authors:  J R Dipersio; S M Ficorilli; F J Varga
Journal:  J Clin Microbiol       Date:  1984-12       Impact factor: 5.948

3.  Clinical impact of rapid identification and susceptibility testing of bacterial blood culture isolates.

Authors:  G M Trenholme; R L Kaplan; P H Karakusis; T Stine; J Fuhrer; W Landau; S Levin
Journal:  J Clin Microbiol       Date:  1989-06       Impact factor: 5.948

  3 in total