| Literature DB >> 7253689 |
J L Myers, A Bull, D G Kastl, W S Pierce.
Abstract
We recently utilized a left ventricular assist device in a patient who could not be separated from cardiopulmonary bypass after aortic and mitral valve replacement. The nonfunctioning pericardial aortic valve became completely occluded owing to a fibrinous fusion of the valve leaflets during the 8-day period of assist pumping. A second operation was required to open the fused valve leaflets. This experience suggests that special care should be taken in selecting an aortic valve prosthesis in a patient who may require the left ventricular assist device in order to be separated from cardiopulmonary bypass.Entities:
Mesh:
Year: 1981 PMID: 7253689
Source DB: PubMed Journal: J Thorac Cardiovasc Surg ISSN: 0022-5223 Impact factor: 5.209