Literature DB >> 7252820

High-performance liquid chromatographic determination of propranolol and 4-hydroxypropranolol in plasma.

M T Rosseel, M G Bogaert.   

Abstract

A high-performance liquid chromatographic method for the simultaneous determination of propranolol and 4-hydroxypropranolol in plasma is presented. The method involves plasma extraction at basic pH with ethyl column, and fluorescence detection. The within-run a reversed-phase column coefficient of variation were 3.0-7.1% for propranolol and 5.8-8.3% for 4-hydroxypropranolol. The day-to-day variations were 4.6 and 8.2% for propranolol and 4-hydroxypropranolol, respectively. The method can detect 1 ng of each compound/ml of plasma.

Entities:  

Mesh:

Substances:

Year:  1981        PMID: 7252820     DOI: 10.1002/jps.2600700631

Source DB:  PubMed          Journal:  J Pharm Sci        ISSN: 0022-3549            Impact factor:   3.534