Abstract
We described potential interference of procainamide with our high pressure liquid chromatography procedure for assaying theophylline (similar to that given by Orcutt et al.). In participating in the AACC-TDM quality-control program, we discovered procainamide interference on a column dedicated to theophylline and acetaminophen serum analyses. Following attempts at rejuvenation, this 7-month-old column was replaced with a new one, which gave base-line separation of theophylline and procainamide with no difficulty. We recommend close attention to the aging characteristics of columns and participation in a quality-control program.Entities:
Mesh:
Substances:
Year: 1981 PMID: 7233485
Source DB: PubMed Journal: Ther Drug Monit ISSN: 0163-4356 Impact factor: 3.681