Literature DB >> 7217262

Determination of vincamine in human plasma using automated high-performance liquid chromatography.

C Dubruc, H Caqueret, G Bianchetti.   

Abstract

A specific and sensitive high-performance liquid chromatographic method for the quantitative determination of vincamine at therapeutic concentrations in plasma is described. The column was packed with Spherisorb ODS 5 micrometer, and the mobile phase was acetonitrile-potassium phosphate (0.02 M, pH 2.3) (50:50) with a flow-rate of 10 ml/min. Detection was at 230 nm. Using automated large-volume injection of a non-eluting solvent (0.02 M potassium phosphate) the method was capable of the analysis of a large number of samples daily. The coefficient of variation of the procedure was 4.8% at a plasma vincamine concentration of 10 ng/ml

Entities:  

Mesh:

Substances:

Year:  1981        PMID: 7217262     DOI: 10.1016/s0021-9673(00)81676-3

Source DB:  PubMed          Journal:  J Chromatogr


  1 in total

1.  Ion-molecule reactions of oxygenated chemical ionization reagents with vincamine.

Authors:  G F Bauerle; B J Hall; N V Tran; J S Brodbelt
Journal:  J Am Soc Mass Spectrom       Date:  1996-03       Impact factor: 3.109