Literature DB >> 718403

[Comparative experimental investigations with bioglass (L. L. Hench) and Al2O3-ceramic coated with mod. bioglass. II. Results of experiments with loaded implants (author's transl)].

P Griss, E Werner, G Heimke, U Raute-Krenisen.   

Abstract

This report is based on 4 comparative experimental series is sheep using a cementfree total hip replacement model. Basically a smooth stem design was compared to a step stem design for biomechanical interlocking anchorage. Both designs were implanted as pure high density Al2O3-ceramic components and as Al2O3-ceramic parts coated with Bioglass or Bioglass-ceramic (post op. follow up 13 months, 17 animals). None of the 4 stem modifications showed a tendency for permanent stabilisation or bonding in bone. Besides insufficient bonding of the glass coatings to the substrate and apparent biodegradability of the bioglass coatings in the body, insufficient biomechanical knowledge of endosteal direct anchorage of prosthetic devices is the main reason for failure in these experiments. An exclusive biochemical way of anchoring prosthetic parts in the marrow cavity is rejected. The results presented are discussed in the light of the recent literature.

Entities:  

Mesh:

Substances:

Year:  1978        PMID: 718403     DOI: 10.1007/bf00397960

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0344-8444


  25 in total

1.  [Tissue adaptability processes involving self-supporting endoprostheses with rib anchors].

Authors:  J Harms; E Mäusle; G Biehl
Journal:  Z Orthop Ihre Grenzgeb       Date:  1975-08

2.  Biological activity and histocompatibility of dense Al 2 O 3 -MgO ceramic implants in rats.

Authors:  P Griss; H V von Andrian-Werburg
Journal:  J Biomed Mater Res       Date:  1973

3.  Attachment of prostheses to the musculoskeletal system by tissue ingrowth and mechanical interlocking.

Authors:  S F Hulbert; F W Cooke; J J Klawitter; R B Leonard; B W Sauer; D D Moyle; H B Skinner
Journal:  J Biomed Mater Res       Date:  1973

4.  Porous implant systems for prosthesis stabilization.

Authors:  C A Homsy; T E Cain; F B Kessler; M S Anderson; J W King
Journal:  Clin Orthop Relat Res       Date:  1972       Impact factor: 4.176

5.  Attachment of bone to threaded implants by ingrowth and mechanical interlocking.

Authors:  P Predecki; B A Auslaender; J E Stephan; V L Mooney; C Stanitski
Journal:  J Biomed Mater Res       Date:  1972-09

6.  [The reaction of bone to bone-cement in the replacement of the hip joint].

Authors:  H G Willert; P Puls
Journal:  Arch Orthop Unfallchir       Date:  1972

7.  Sintered fiber metal composites as a basis for attachment of implants to bone.

Authors:  J Galante; W Rostoker; R Lueck; R D Ray
Journal:  J Bone Joint Surg Am       Date:  1971-01       Impact factor: 5.284

8.  [The Mannheimer oxide ceramic-metal composite hip prostheses (author's transl)].

Authors:  P Griss; E Werner; R Buchinger; G Heimke
Journal:  Arch Orthop Unfallchir       Date:  1977-01-21

9.  Analysis of bioglass fixation of hip prostheses.

Authors:  L L Hench; C G Pantano; P J Buscemi; D C Greenspan
Journal:  J Biomed Mater Res       Date:  1977-03

10.  [Comparative experimental investigations with bioglass (L. L. Hench), Al2O3-ceramic and Al2O3-ceramic coated with a mod. Bioglass. I. Results of experiments under non-loaded conditions (author's transl)].

Authors:  P Griss; E Werner; G Heimke; R Buchinger
Journal:  Arch Orthop Unfallchir       Date:  1977-11-21
View more
  1 in total

1.  Ceramic implant materials.

Authors:  G Heimke; P Griss
Journal:  Med Biol Eng Comput       Date:  1980-07       Impact factor: 2.602