| Literature DB >> 7161377 |
Abstract
Replicate control tests were performed with the agar overlay and Kirby-Bauer techniques, using two sources of Mueller-Hinton agar. Mean zones observed with the two methods differed by 1 to 3 mm. Despite the minor differences between methods, National Committee for Clinical Laboratory Standards quality control limits which were established for the Kirby-Bauer method could also be used for the agar overlay technique.Entities:
Mesh:
Substances:
Year: 1982 PMID: 7161377 PMCID: PMC272554 DOI: 10.1128/jcm.16.6.1145-1147.1982
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948