In vitro closing behaviour of Björk-Shiley, St Jude and Hancock heart valve prostheses in relation to the in vivo recorded aortic valve closure.

A A van Steenhoven, T J van Duppen, J W Cauwenberg, R J van Renterghem.   

Abstract

To compare prosthetic valve behaviour with that of the natural aortic valve, experiments were performed in vitro as well as in vivo. In a mock circulation system, cinematographic high-speed recordings of the valvular behaviour were made for Björk-Shiley, St Jude and Hancock heart valve prostheses. Simultaneously with the film recording, the aortic flow and the left ventricular and aortic pressures were measured. The closing behaviour of the natural aortic valve was recorded in in vivo experiments following the same measuring technique. Comparison of the film frames with the aortic flow signal revealed that the mechanical prostheses mainly close due to the back flow in the early phase of diastole; they close only for 5% of their cross-sectional area during systolic ejection. The Hancock bioprosthesis closes already for 45% during the flow deceleration phase of systole, which is however significantly less than the corresponding closure of the aortic valve as recorded in vivo (74%). The change in fluid viscosity from 3.10(-3) Ns/m2 to 1.10(-3) Ns/m2 does not affect the closing behaviour of the mechanical prostheses. Reducing the peak value of systolic aortic flow by a factor two decreases the cross-sectional area of the prosthetic valves at peak systole. When the natural aortic valve closing mechanism is applied to a disc-valve prosthesis, a partial (30%) systolic valve closure is found.

Entities:  

Mesh:

Year:  1982        PMID: 7161286     DOI: 10.1016/0021-9290(82)90049-5

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  2 in total

1.  Thrombogenic potential of transcatheter aortic valve implantation with trivial paravalvular leakage.

Authors:  Lawrence N Scotten; Rolland Siegel
Journal:  Ann Transl Med       Date:  2014-05

2.  Dynamics of a mechanical monoleaflet heart valve prosthesis in the closing phase: effect of squeeze film.

Authors:  C Gill-Jeong; K B Chandran
Journal:  Ann Biomed Eng       Date:  1995 Mar-Apr       Impact factor: 3.934

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.