Evaluation of a dynamometer for measurement of isometric and isokinetic torques.

M Ericsson, K Johansson, B Nordgren, L O Nordesjö, O Borges.   

Abstract

A modified version of the original Cybex II dynamometer with a temperature compensated strain gauge transducer has been evaluated. It was found that the calibration constant was dependent of the vertical load of the device. The ratio (k) between the applied and the recorded torque is angle independent during extension and k is equal to 1.18 and during flexion k is angle dependent and can be expressed as k = -.0008 theta + 1.208, where theta is the angle of flexion.

Mesh:

Year:  1982        PMID: 7157555     DOI: 10.3109/03009738209178427

Source DB:  PubMed          Journal:  Ups J Med Sci        ISSN: 0300-9734            Impact factor:   2.384