| Literature DB >> 7125962 |
R Kiraly, R Yozu, D Hillegass, H Harasaki, S Murabayashi, J Snow, Y Nosé.
Abstract
A trileaflet valve, compression molded of Hexsyn rubber, was evaluated as a component in temporary blood pumps. A pair of valves was integrated into an LVAD which was tested in vitro and in vivo. The valve's hemodynamic performance was satisfactory and the blood compatibility was excellent in a three-week, in vivo experiment conducted to demonstrate performance in a short-term application. Hexsyn, which has excellent properties and produces high quality, complex components with ease of fabrication, offers a potential for low-cost blood pump valves.Mesh:
Substances:
Year: 1982 PMID: 7125962 DOI: 10.1111/j.1525-1594.1982.tb04082.x