Literature DB >> 7118644

Measurement of gas viscosity with a Fleisch pneumotachograph.

D P Johns, J J Pretto, J A Streeton.   

Abstract

A simple rapid method of measuring gas viscosity using a standard Fleisch pneumotachograph and a 3-liter hand-driven syringe is described. Comparison of pneumotachographic and predicted viscosity of five pure gases (CO2, N2, He, O2, and Ar), two binary mixtures (He-O2 and N2-O2), and one quaternary mixture (He-air) gave an overall coefficient of correlation of 0.987 and an accuracy of better than 1.7%. Our data show the well-known marked curvilinear relationship between viscosity and the concentration of helium in air and oxygen mixtures. These studies indicate that a Fleisch pneumotachograph can be used as a simple accurate gas viscometer to characterize gas mixtures in terms of viscosity.

Entities:  

Mesh:

Substances:

Year:  1982        PMID: 7118644     DOI: 10.1152/jappl.1982.53.1.290

Source DB:  PubMed          Journal:  J Appl Physiol Respir Environ Exerc Physiol        ISSN: 0161-7567


  2 in total

1.  Calibration of Fleisch and screen pneumotachographs for use with various oxygen concentrations.

Authors:  M J Turner; I M MacLeod; A D Rothberg
Journal:  Med Biol Eng Comput       Date:  1990-03       Impact factor: 2.602

2.  Evaluation of a Pitot type spirometer in helium/oxygen mixtures.

Authors:  S Søndergaard; S Kárason; S Lundin; O Stenqvist
Journal:  J Clin Monit Comput       Date:  1998-08       Impact factor: 2.502