A double beam Doppler ultrasound method for quantitative blood flow velocity measurement.

W Wei-qi, Y Lin-xin.   

Abstract

A method for measuring the absolute blood flow velocity waveform is reported. Two independent beams of ultrasound illuminate a vessel simultaneously, producing complementary Doppler signals. The two Doppler frequency shift signals are processed by subtraction and addition at the receiver. The optimum probe position where blood flow velocity is detected can be found as the position where the subtractor output reaches zero. At this position the blood flow velocity is the output from the adder. By this means the influence of the angle between the probe and blood flow is eliminated so that a quantitative measurement is obtained. Both in vitro and clinical results are reported.

Entities:  

Mesh:

Year:  1982        PMID: 7112728     DOI: 10.1016/s0301-5629(82)80010-0