| Literature DB >> 709821 |
M G DeRuyter, A P De Leenheer.
Abstract
We propose a single-run liquid-chromatographic determination, with ultraviolet detection at 330 nm, for serum retinol and retinyl esters. The vitamin A derivatives are extracted according to the Bligh-Dyer procedure. With 200 microliter or serum, the lower detection limit is 50 microgram/liter for retinol and about 100 microgram/liter for retinyl esters. Within-run precision (CV) was 2.3% for retinol, 4.3% for retinyl palmitate. Day-to-day percision (CV, n = 20) for retinol was 4.9% during a month. The method can be used for the assessment of vitamin A absorption tests and for the determination of serum retinol (normal, subnormal, and above-normal concentrations). Serum retinyl esters can only be measured in conditions where concentrations exceed 100 microgram/liter.Entities:
Mesh:
Substances:
Year: 1978 PMID: 709821
Source DB: PubMed Journal: Clin Chem ISSN: 0009-9147 Impact factor: 8.327