| Literature DB >> 7094922 |
K S Harrison, X Liu, S T Han, E E Camargo, H N Wagner.
Abstract
A miniature CdTe probe interfaced to a microcomputer was used to measure left ventricular ejection fraction (LVEF) in 25 patients. LVEF obtained with the CdTe module, in the beat-to-beat mode, or the integrated gated mode agreed well with LVEF obtained with a gamma camera (r = 0.80; r = 0.82 respectively). Similarly, LVEF by CdTe probe agreed with LVEF obtained by gated equilibrium studies performed with a computerized NaI probe. The CdTe probe can provide comparable measurement of LVEF at a fraction of the cost of a camera-computer system and, being small and lightweight, the CdTe probe is adaptable for monitoring patients in intensive care facilities.Entities:
Mesh:
Substances:
Year: 1982 PMID: 7094922 DOI: 10.1007/BF00256464
Source DB: PubMed Journal: Eur J Nucl Med ISSN: 0340-6997