| Literature DB >> 7091278 |
Abstract
We compared the maximum light output of selected 19- and 20-gauge continuous-fiber and detachable-tip endoilluminators to the retinal damage threshold for intravitreal fiber-optic light in the owl monkey. Because of cumulative light toxicity to the retina there is a substantial risk of retinal damage from endoillumination when the endoilluminator is close to the retina (for example, during membrane peeling procedures) and especially when the continuous-fiber endoilluminator with high light output is used.Entities:
Mesh:
Year: 1982 PMID: 7091278 DOI: 10.1016/0002-9394(82)90186-6
Source DB: PubMed Journal: Am J Ophthalmol ISSN: 0002-9394 Impact factor: 5.258