| Literature DB >> 7085055 |
D T Shakespeare, N J Henderson, G Clough.
Abstract
The slit catheter, a new instrument for the measurement of compartment pressure, has been compared with the wick catheter in experimental conditions. The two catheters were inserted into the anterior tibial compartments of 8 human legs and controlled external pressures were applied by means of an inflatable limb bag. The mean resting intramuscular pressure of 8.5 +/- 6.2 mmHg for the slit and 8.7 mmHg +/- 6.2 mmHg for the wick catheter agrees with that of other workers, and the slit catheter records a pressure at all applied external pressures not significantly different from the wick catheter (paired t test). The slit catheter provides a simple and inexpensive way of estimating compartment pressure which is as accurate as a method widely used both experimentally and clinically in North America.Entities:
Mesh:
Year: 1982 PMID: 7085055 DOI: 10.1016/0020-1383(82)90094-8
Source DB: PubMed Journal: Injury ISSN: 0020-1383 Impact factor: 2.586