C E Hahn, E A Hall, P Maynard, W J Albery.
Abstract
A prototype composite electrode consisting of a metallized membrane "outer" cathode, and a conventional "inner" silver disc cathode, both sharing the same electrolyte, is described. Preliminary results indicate that the metallized membrane will act as an oxygen filter, allowing anaesthetic gases such as nitrous oxide, to pass through to the second (inner) cathode, where they may be reduced. Linear relationships exist for oxygen concentration and metallized membrane current; and for nitrous oxide concentration and the second cathodic current. Extension of this technique to measure both halothane and nitrous oxide on the second cathode, simultaneously with oxygen on the metallized membrane, is discussed.Entities:
Mesh:
Substances:
Year: 1982 PMID: 7082528 DOI: 10.1093/bja/54.6.681