D A Rottenberg, H C Lu, K J Kearfott.
Abstract
We have studied the effect of brain tissue (gray matter-white matter) heterogeneity and computerized tomography (CT) noise on the accuracy of xenon CT measurements of regional cerebral blood flow (rCBF) based upon "autoradiographic" and multiple-scan washin protocols. The results of our mathematical analysis indicate that both protocols are associated with a variety of measurement errors that lead to unacceptable and, to a large extent, unpredictable uncertainties in calculated values of rCBF. Brain tissue heterogeneity and high volumetric flow rates may--even in the absence of movement artifact and CT noise--lead to measurement errors in excess of 20%. Moreover, CT noise is additive in regard to these errors, and constitutes the most confounding variable of all.Entities:
Mesh:
Substances:
Year: 1982 PMID: 7076729 DOI: 10.1038/jcbfm.1982.16
Source DB: PubMed Journal: J Cereb Blood Flow Metab ISSN: 0271-678X Impact factor: 6.200