| Literature DB >> 707407 |
Abstract
A disposable bleeding time device that provides two simultaneous standardized incisions is described. The mean bleeding time of 47 normal adults was 4.1 min with a 95% range of 2.2--7.0 min. The standard deviation of duplicate bleeding times was 0.7 min, and the day-to-day standard deviation for individuals was 0.9 min. In a double-blind crossover study of 20 normal adults, the mean bleeding time increased from 3.7 to 6.2 min after ingestion of 1 g aspirin. The device is extremely simple to use and is essentially painless. Physical trauma is minimal.Entities:
Mesh:
Substances:
Year: 1978 PMID: 707407 DOI: 10.1093/ajcp/70.3.406
Source DB: PubMed Journal: Am J Clin Pathol ISSN: 0002-9173 Impact factor: 2.493