Physiologic implications of two different heat shields for infants under radiant warmers.

S Baumgart, W W Fox, R A Polin.   

Abstract

To study the effects of two different heat shields commonly used under radiant warmers (thin plastic blanket vs a plastic body hood) on premature newborn infants, eight neonates were studied to measure insensible water loss and radiant power density at the skin under control and two shielded conditions. The body hood was found to diminish transmission of radiant heat to the infant's skin by 80%, absorbing heat and becoming warm (P less than 0.001). The plastic blanket absorbed only 15% of radiant warmer heat output (P less than 0.01). Insensible water loss was significantly less under the plastic blanket (1.88 +/- 0.4) ml/kg/hour) than under control (2.70 +/- 0.50 ml/kg/hour, P less than 0.01) and hood (2.86 +/- 0.32 ml/kg/hour, P less than 0.05) conditions. There was no decrease in insensible water loss under the hood compared to controls. This study demonstrates that a thin plastic blanket is the more effective shield against insensible water loss under a radiant warmer. Caution should be exercised with any shielding device to prevent interference with radiant heat delivery.

Entities:  

Mesh:

Year:  1982        PMID: 7069544     DOI: 10.1016/s0022-3476(82)80594-5

Source DB:  PubMed          Journal:  J Pediatr        ISSN: 0022-3476            Impact factor:   4.406


  3 in total

Review 1.  Renal failure in the newly born.

Authors:  J T Brocklebank
Journal:  Arch Dis Child       Date:  1988-08       Impact factor: 3.791

2.  Open or closed incubators.

Authors:  D Hull; A Wheldon
Journal:  Arch Dis Child       Date:  1986-02       Impact factor: 3.791

3.  Phototherapy and the use of heat shields in very low birthweight infants.

Authors:  P R Stutchfield; N Modi; A M Weindling
Journal:  Arch Dis Child       Date:  1988-05       Impact factor: 3.791

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.