Clinical experience with the Fieldmaster perimeter.

J C Bobrow, R C Drews.   

Abstract

Perimetry performed on a Fieldmaster Model 200 on a consecutive series of 167 patients in a clinical practice setting demonstrated that Fieldmaster testing elicited 95% of defects found on Goldmann kinetic perimetry with very few false-positive or false-negative results; relatively unskilled personnel were able to conduct Fieldmaster testing well; and as our personnel and patients became familiar with the test, the rate of incomplete examinations declined to less than 2%.

Entities:  

Mesh:

Year:  1982        PMID: 7065096     DOI: 10.1016/0002-9394(82)90420-2

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258