Comparison of oculoplethysmography/carotid phonoangiography with duplex scan/spectral analysis in the detection of carotid artery stenosis.

B A Keagy, W F Pharr, D Thomas, D E Bowes.   

Abstract

The accuracy of the duplex scan with spectral analysis (DS/SA) in predicting the presence of arteriographic carotid stenosis was compared to that of oculoplethysmography/carotid phonoangiography (OPG/CPA) in 234 vessels from 117 patients who had had both non-invasive studies in addition to independently interpreted arteriograms. The DS/SA with 212/234 (91%) overall correct responses was superior to the OPG/CPA which properly classified 181/234 (77%) of the vessels (p less than .01). Of major clinical impact was the superiority of the DS/SA (p less than .001) in identifying the 72 vessels with 50-99% stenosis. The OPG/CPA had a discouraging 39/72 (54%) false-negative rate in this group whereas the DS/SA missed only 9/72 (12%) of these arteries. Of those 39 incorrect responses for the OPG/CPA, 26/39 (67%) were in patients with a 50% or greater stenosis on the contralateral side. This is a recognized area of weakness for that test. Based on the results of this study, we have abandoned the use of the OPG/CPA in the evaluation of patients with suspected carotid stenosis and rely solely on the duplex scan with spectral analysis.

Entities:  

Mesh:

Year:  1982        PMID: 7064179     DOI: 10.1161/01.str.13.1.43

Source DB:  PubMed          Journal:  Stroke        ISSN: 0039-2499            Impact factor:   7.914


  2 in total

1.  Duplex carotid ultrasound.

Authors:  J D Acker; C A Cole; K A Mauney; D M Joly; J E Machin
Journal:  Neuroradiology       Date:  1986       Impact factor: 2.804

2.  Carotid Doppler evaluation in cerebrovascular disease.

Authors:  J G D'Alton; J W Norris
Journal:  Can Med Assoc J       Date:  1983-12-01       Impact factor: 8.262

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.