Literature DB >> 7046332

A portable PEEP valve for 0-20 cm H2O.

M P Boidin.   

Abstract

Positive end expiratory pressure (PEEP) ventilation is a valuable method in modern critical care therapy. Nevertheless it is often necessary to take patients off the ventilator for routine physiotherapy as well as for special investigations and transport. Portable ventilators or manual bag ventilators are used during the offventilator periods. In order to keep the expiratory pressure in the therapeutic range it is desirable to have a portable PEEP valve available. This valve has to be lightweight, small, easy to use, stable and cheap. AMBU Danmark has designed such a PEEP valve which fits onto adult and pediatric Non-Return (NR) valves. The latest design is a green PVC-spring type valve with a pressure range of 0-20 cm H2O. We have evaluated this valve in the laboratory for different flows, temperature, humidity and position, and in the clinic in various situations. We found the AMBU PEEP valve to be accurate and easy to use with a stable performance meeting all requirements for emergency medical care (EMC).

Entities:  

Mesh:

Year:  1982        PMID: 7046332

Source DB:  PubMed          Journal:  Acta Anaesthesiol Belg        ISSN: 0001-5164


  1 in total

1.  Hemodynamic variations in arterial wave reflection associated with the application of increasing levels of PEEP in healthy subjects.

Authors:  Jacopo Belfiore; Etrusca Brogi; Niccolo Nicolini; Davide Deffenu; Francesco Forfori; Carlo Palombo
Journal:  Sci Rep       Date:  2022-02-28       Impact factor: 4.379

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.