Literature DB >> 7018433

Extrinsic obstruction of the Björk-Shiley valve in the mitral position.

D B Williams, J R Pluth, T A Orszulak.   

Abstract

Four recently seen patients with acute, extrinsic obstruction of a Björk-Shiley prosthesis in the mitral position are described. In 2 of them, the disc occluder was stuck in the closed position by a strand of chordal tissue, which was compressed between the disc edge and valve rim. In another patient, sticking of the disc by a similar mechanism was caused by a suture that migrated into the valve orifice. Intermittent malfunction of the valve in the fourth patient was related to interference with valve opening by ventricular myocardium and occurred when the patient lay on his right side. The mechanism of extrinsic obstruction of the Björk-Shiley valve and their recognition, treatment, and prevention are discussed.

Entities:  

Mesh:

Year:  1981        PMID: 7018433     DOI: 10.1016/s0003-4975(10)61375-6

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

1.  Obstruction of a Björk-Shiley mitral prosthesis in a patient with idiopathic hypertrophic subaortic stenosis.

Authors:  T Raychaudhury; A Faichney; P R Walbaum
Journal:  Thorax       Date:  1983-05       Impact factor: 9.139

2.  Intraoperative detection of stuck leaflet of prosthetic mitral valve.

Authors:  Monish S Raut; Arun Maheshwari; Sumir Dubey
Journal:  Indian Heart J       Date:  2017-06-04
  2 in total