Literature DB >> 6995495

An evaluation of the Replireader in the identification of Enterobacteriaceae isolated from urine and in the recording of sensitivity tests performed by an agar dilution method.

P M Waterworth.   

Abstract

In the Replireader system for identifying Enterobacteriaceae, plates of biochemical media are inoculated with a replicator and the results are put into a computer. The machine correctly identified 92x2% of 734 strains of Gram-negative bacilli isolated from urine; it was incorrect in 0x8% and failed to recognise 7%. The Replireader was also used to record the results of sensitivity tests using a plate dilution method in which the drugs were provided by impregnated filter papers (Adapads).

Mesh:

Year:  1980        PMID: 6995495      PMCID: PMC1146145          DOI: 10.1136/jcp.33.6.571

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  1 in total

1.  Evaluation of the Repliscan system for identification of Enterobacteriaceae.

Authors:  S D Brown; J A Washington
Journal:  J Clin Microbiol       Date:  1978-12       Impact factor: 5.948

  1 in total