Determination of methotrexate in serum by a rapid, fully mechanized enzyme immunoassay (EMIT).

M Oellerich, P Engelhardt, M Schaadt, V Diehl.   

Abstract

A homogeneous enzyme immunoassay for the determination of methotrexate in serum (EMIT, Syva Corp.) was fully mechanized by the use of an Eppendorf analyzer 5010. Coefficients of variation from day to day were in the range 2--16%. A comparison of the results obtained by EMIT and a radioimmunoassay (Methotrexate 125 I Radioimmunoassay Kit, Diagnostic Biochemistry Inc.) in a series of 50 samples from patients showed a good correlation between both methods. The specificity of the EMIT assay appeared to be adequate. The time required for the determination of about 20 patient samples by this procedure was only about 80 minutes. From our clinical experimence the EMIT assay appears to be well suited for routine monitoring of methotrexate serum concentrations above 0.1 mumol/1 during intrathecal and high dose methotrexate therapy.

Entities:  

Mesh:

Substances:

Year:  1980        PMID: 6991631     DOI: 10.1515/cclm.1980.18.3.169

Source DB:  PubMed          Journal:  J Clin Chem Clin Biochem        ISSN: 0340-076X


  2 in total

1.  Pharmacokinetic study of methotrexate, folinic acid and their serum metabolites in children treated with high-dose methotrexate and leucovorin rescue.

Authors:  C Wolfrom; R Hepp; R Hartmann; H Breithaupt; G Henze
Journal:  Eur J Clin Pharmacol       Date:  1990       Impact factor: 2.953

2.  Establishing Syva Methotrexate Assay on Siemens Dimension RxL Analyzer: Experience in a Tertiary Cancer Care Laboratory.

Authors:  Preeti Chavan; Vivek Bhat; Ashwadeep Karmore; Rajani Mohite; Swati Waykar; Hemali Kadu
Journal:  J Lab Physicians       Date:  2017 Jan-Mar
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.