E J Ruitenberg, V M Sekhuis, B J Brosi.
Abstract
The performance of a prototype of a now commercially available multiple-channel photometer for high-speed reading of microplates to be used in the enzyme-linked immunosorbent assay was evaluated. For this purpose the extinction of a stable test liquid dispensed in flat-bottomed polystyrene microplates was measured, varying the test volume and the optical density. From the results obtained for dilutions of a single, biolgoically inactive material we concluded that the photometric inaccuracy is negligible when performing enzyme-linked immunosorbent assay with biologically active samples.Entities: