G M Deeb, H S Borovetz, B P Griffith, R L Hardesty.
Abstract
The dilution technique for determining cardiac output using indocyanine green dye is limited in patients weighing less than 20 kg because of the obligatory volume loss. Reproducible achieved using the green dye dilution method by the establishment of a low flow peripheral arteriovenous shunt. The shunt materials were treated with thromboresistant agents--TDMAC (7%) and albumin (1 g/dl)--to facilitate the use of this technique without heparin. For A-V shunt flow rates of 8-30 cc/min reproducible values of cardiac output were obtained for up to 38 hours which were in good agreement with determinations made using the conventional technique of dye dilution.Entities:
Mesh:
Substances:
Year: 1980 PMID: 6986120 PMCID: PMC1344627 DOI: 10.1097/00000658-198001000-00020
Source DB: PubMed Journal: Ann Surg ISSN: 0003-4932 Impact factor: 12.969