Literature DB >> 6985851

A comparison of serum phenytoin determination by the substrate-labeled fluorescent immunoassay with gas chromatography, liquid chromatography, radioimmunoassay and "EMIT".

R C Wong, R George, R Yeung, J F Burd.   

Abstract

Patients' sera were analyzed for phenytoin by the Substrate-Labeled Fluorescent Immunoassay (SLFIA), gas chromatography (GC), liquid chromatography (LC), Radioimmunoassay (RIA) and EMIT and the results were compared. The EMIT assay was performed using a "PACER Analyzer". All assay results compared favorably. The correlation coefficients were: SLFIA vs GC, 0.995 (n = 45); SLFIA vs. LC, 0.993 (n = 37); SLFIA vs. EMIT, 0.965 (n = 67); SLFIA vs. RIA, 0.977 (n = 34). These results show that phenytoin levels determined by the SLFIA compare well with those obtained by the other four assay techniques.

Entities:  

Mesh:

Substances:

Year:  1980        PMID: 6985851     DOI: 10.1016/0009-8981(80)90187-4

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786