Compliance of the tubing compartment of lung ventilators.

G Verheecke, A Gilbertson.   

Abstract

The compliance of three types of breathing tubing commonly used in lung ventilators was measured. The black rubber corrugated tubing commonly used in anaesthetic circuits is too compliant for use in ventilators in intensive care. It is suggested that the international draft standard of compliance for breathing tubes used in lung ventilators should be amended.

Mesh:

Year:  1981        PMID: 6948881     DOI: 10.1007/bf01709728

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  1 in total

1.  The internal compliance of ventilators.

Authors:  L Loh; M K Chakrabarti
Journal:  Anaesthesia       Date:  1971-10       Impact factor: 6.955

  1 in total
  2 in total

1.  Positive end-expiratory pressure: implications for tidal volume changes in anesthesia machine ventilation.

Authors:  W R Elliott; A E Harris; J H Philip
Journal:  J Clin Monit       Date:  1989-04

2.  Exhalatory dynamic interactions between patients connected to a shared ventilation device.

Authors:  Pedro M Garcia Eijo; Juan D'Adamo; Arturo Bianchetti; Thomas Duriez; Juan M Cabaleiro; Célica Irrazabal; Pablo Otero; Guillermo Artana
Journal:  PLoS One       Date:  2021-05-04       Impact factor: 3.240