| Literature DB >> 6948881 |
Abstract
The compliance of three types of breathing tubing commonly used in lung ventilators was measured. The black rubber corrugated tubing commonly used in anaesthetic circuits is too compliant for use in ventilators in intensive care. It is suggested that the international draft standard of compliance for breathing tubes used in lung ventilators should be amended.Mesh:
Year: 1981 PMID: 6948881 DOI: 10.1007/bf01709728
Source DB: PubMed Journal: Intensive Care Med ISSN: 0342-4642 Impact factor: 17.440