A quantitative technical quality assay method for electrocardiograms.

L T Sheffield, D I Roitman, S Kansal.   

Abstract

A method is presented for quantitative assessment of ECG technical quality to permit sensitive detection of performance trends of individual technicians or entire ECG laboratories. The common flaws in ECG recording technic are those which introduce ultra-low frequency artifact or baseline shift, powerline artifact, myographic and intermediate frequency artifact, clipping of waveforms, disconnection or reversal of electrodes and general faults of protocol adherence. Procedures for quantitating these attributes are presented and explained in order to arrive at a mean figure of merit for each ECG recording. Use of such a method can aid ECG technicians' understanding of the important features of tracing quality and provide an objective basis for the recognition of outstanding performance, on one hand, or the need for remedial training, on the other.