Literature DB >> 6884124

Testing the hydrocephalus shunt valve.

C Watts, H D Keith.   

Abstract

Unwanted results of shunting for hydrocephalus include slit and asymmetrical ventricles and the conversion of communicating hydrocephalus to noncommunicating hydrocephalus. Evidence suggests this may be due to a mismatch between the pressure/flow performance characteristics of the shunt valve and the pathologic hydrodynamics of the ventricular cerebrospinal fluid. Commercially available shunts are currently tested using a steady-state pressure/flow method. We have utilized a bench testing method which incorporates pulsed flow and varying compliance. Results indicate that with minimal compliance in the system, opening and closing pressures differ markedly in value with pulsed flow as compared to steady-state testing. The greater the compliance the less deviation was noted. It is recommended more effort be made to understand the value of matching pressure/flow characteristics of valve to the individual hydrocephalic state.

Entities:  

Mesh:

Year:  1983        PMID: 6884124     DOI: 10.1159/000120117

Source DB:  PubMed          Journal:  Childs Brain        ISSN: 0302-2803


  2 in total

1.  Rationale and methodology of the multicenter pediatric cerebrospinal fluid shunt design trial. Pediatric Hydrocephalus Treatment Evaluation Group.

Authors:  J M Drake; J Kestle
Journal:  Childs Nerv Syst       Date:  1996-08       Impact factor: 1.475

2.  Cerebrospinal fluid shunt dynamics in patients with idiopathic adult hydrocephalus syndrome.

Authors:  J Malm; B Kristensen; M Fagerlund; L O Koskinen; J Ekstedt
Journal:  J Neurol Neurosurg Psychiatry       Date:  1995-06       Impact factor: 10.154

  2 in total