Multicentre comparison of nine coagulometers and manual tilt-tube methods for prothrombin time performance.

.   

Abstract

This multicentre study compares in parallel nine commercial coagulometers and manual tilt-tube methods for their capacity to distinguish two pairs of lyophilized plasmas with slightly different prothrombin times (PTs). The study was based on the rationale that the more precise a method is, the better it is able to discriminate between plasmas with similar PTs. Good or excellent discrimination, as judged by higher Student t values, were achieved slightly more frequently by coagulometers than by manual methods, particularly when paris of plasmas with smaller PT differences were tested.

Mesh:

Substances:

Year:  1983        PMID: 6883969     DOI: 10.1111/j.1365-2257.1983.tb01350.x

Source DB:  PubMed          Journal:  Clin Lab Haematol        ISSN: 0141-9854