Literature DB >> 6872210

Potentiometry of Na+ in undiluted serum and urine with use of an improved neutral carrier-based solvent polymeric membrane electrode.

P Anker, H B Jenny, U Wuthier, R Asper, D Ammann, W Simon.   

Abstract

We present an improved Na+-selective liquid membrane electrode for measurement of Na+ concentrations in both undiluted serum and urine. The values for urinary Na+ obtained with the ion-selective electrode agree well with those obtained with the flame photometer. The correlation gives a standard residual deviation of +/- 2.7 mmol/L over the Na+ range of 25-280 mmol/L. In serum, this direct potentiometry yields Na+ concentrations 5.4% (SD 1.1%) higher than those obtained by atomic spectrometry and a standard residual deviation of +/- 1.1 mmol/L. Correction of these values for the volumes of protein and lipid leads to potentiometric values 1.2% (SD 0.7%) lower than those by flame photometry (residual standard deviation: +/- 1.0 mmol/L). Other factors that possibly contribute to this discrepancy are discussed.

Entities:  

Mesh:

Substances:

Year:  1983        PMID: 6872210

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  1 in total

1.  New bis(azobenzocrown)s with dodecylmethylmalonyl linkers as ionophores for sodium selective potentiometric sensors.

Authors:  Elżbieta Luboch; Maciej Jeszke; Mirosław Szarmach; Natalia Łukasik
Journal:  J Incl Phenom Macrocycl Chem       Date:  2016-09-29       Impact factor: 1.633