Literature DB >> 68628

[Disorders of blood coagulatiion during haemoperfusion (author's transl)].

H Niessner, M Pichler.   

Abstract

Charcoal haemoperfusion of patient may cause a fall in thrombocyte and fibrinogen levels. The possible cause of these changes in haemostasis was investigated by means of five haemoperfusion devices which were filled with 2000 ml ACD fresh blood and perfused for three hours. A continous reduction in the thrombocytes and the plasma coagulation proteins resulted, while the ethanol test and the level of fibrin degradation products stayed within normal levels. These results indicate the adsorption of thrombocytes and protein on to the surface of the charcoal haemoperfusion device. An activation of the coagulation system (i. e. disseminated intravasal coagulation, fibrinolysis) appears unlikely.

Entities:  

Mesh:

Substances:

Year:  1977        PMID: 68628

Source DB:  PubMed          Journal:  Wien Klin Wochenschr        ISSN: 0043-5325            Impact factor:   1.704


  2 in total

1.  The influence of haemoperfusion on haemostasis and cellular constituents of the blood in the treatment of intoxications: a comparative study of three types of columns (Haemocol, Amberlite XAD-4, Gambro Adsorba 300 C).

Authors:  B Sangster; A N van Heijst; J J Sixma
Journal:  Arch Toxicol       Date:  1981-07       Impact factor: 5.153

2.  The influence of haemoperfusion on white cells, immunoglobulin concentrations and the complement system in the treatment of intoxications.

Authors:  J H Rommes; B Sangster; F H Gmelig Meyling; M R Daha; A N van Heijst
Journal:  Arch Toxicol       Date:  1983-02       Impact factor: 5.153

  2 in total