Literature DB >> 6853274

Fletcher-Suit-Delclos gynecologic applicator: evaluation of a new instrument.

J S Haas, R D Dean, C M Mansfield.   

Abstract

A new instrument has been added to the Fletcher-Suit-Delclos group of gynecologic applicators. The colpostats can be used as mini-ovoids, but by adding a shielded cap, the instrument has the configuration of the original preload Fletcher colpostat. Dosimetric studies show that this applicator produced transmission ratios (the fraction of radiation transmitted through the tungsten shield) and isodose curves similar to the bladder trigone and anterior rectal wall of 10-25%. With the shield-containing cap removed, the mini-ovoid provides little reduction in dose to those areas and should be used with caution.

Mesh:

Year:  1983        PMID: 6853274     DOI: 10.1016/0360-3016(83)90246-8

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  2 in total

1.  Commissioning a CT-compatible LDR tandem and ovoid applicator using Monte Carlo calculation and 3D dosimetry.

Authors:  Justus Adamson; Joseph Newton; Yun Yang; Beverly Steffey; Jing Cai; John Adamovics; Mark Oldham; Junzo Chino; Oana Craciunescu
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

Review 2.  Evolution of Brachytherapy Applicators for the Treatment of Cervical Cancer.

Authors:  Ankur Mourya; Lalit Mohan Aggarwal; Sunil Choudhary
Journal:  J Med Phys       Date:  2021-12-31
  2 in total