Literature DB >> 6848284

Measurement of glucose in plasma by a differential pH technique.

M Luzzana, G Dossi, A Mosca, A Granelli, D Berger, E Rovida, M Ripamonti, A Musetti, L Rossi-Bernardi.   

Abstract

A new automatic apparatus based on the differential measurement of pH between two solutions has been developed. Two 25-microL (internal volume) glass capillary electrodes are used to measure the results of automated (under microcomputer control) chemical reactions that lead to the liberation or the uptake of hydrogen ions. The sensitivity of the differential pH measurements is better than +/- 0.0001 pH unit, and the change in H+ concentration that can be detected by such an apparatus is 1 mumol/L for plasma and 3 mumol/L for whole blood. The technique has been applied to the measurement of glucose in plasma, giving results in agreement with the specifications of the Food and Drug Administration reference method for quantitative determination of glucose (hexokinase/glucose-6-phosphate dehydrogenase method).

Entities:  

Mesh:

Substances:

Year:  1983        PMID: 6848284

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  1 in total

1.  Whole blood L-lactate assay by a new differential pH method: application to metabolic investigations.

Authors:  L Luzi; M Ripamonti; C Marconi; E Cazzola; A Mosca
Journal:  Acta Diabetol Lat       Date:  1990 Apr-Jun