An evaluation of the performance characteristics of different types of collimators used with the EMI brain scanner (MKI) and their significance in specific clinical applications.

S R Thomas, A J Schneider, J G Kereiakes, R R Lukin, A A Chambers, T A Tomsick.   

Abstract

The characteristics of various types of collimators have been studied systematically for the EMI Mark I brain scanner. The degree of collimation ranged from the 13-mm adjacent A and B slices to 3-mm separated slices. Multiple phantom studies (including variable density immiscible liquid interfaces) have been performed to evaluate the effect of collimation on (a) the line-spread function, (b) the change in effective absorption number, and (c) volume averaging. Effects associated with the technique settings involving the kVp, the mA, and the scan time were also investigated. The results of TLD dose measurements as a function of collimation are presented. The significance of collimation in specific clinical studies involving small, fairly well-circumscribed lesions is discussed.

Mesh:

Year:  1978        PMID: 683151     DOI: 10.1118/1.594419

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

Review 1.  Anniversary paper. Development of x-ray computed tomography: the role of medical physics and AAPM from the 1970s to present.

Authors:  Xiaochuan Pan; Jeffrey Siewerdsen; Patrick J La Riviere; Willi A Kalender
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

2.  Volume determinations of the encephalic ventricles with CT and MRI.

Authors:  G D Cramer; D I Allen; L J DiDio
Journal:  Surg Radiol Anat       Date:  1990       Impact factor: 1.246

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.