| Literature DB >> 6816808 |
Abstract
Three techniques are compared for the quantitation of various radiolabeled compounds eluting in the high performance liquid chromatography system. The first technique requires fraction-collecting the effluent from the HPLC, removing an aliquot to scintillation vials, and counting each fraction in a liquid scintillation counter. The second uses direct interface of the HPLC effluent to a flow-through radioactivity detector. The third involves quantitation of various radiolabeled compounds (proteins, steroids, and nucleotides) by splitting the effluent from the HPLC with an electronic steam splitter, thus diverting a present portion to the fraction collector for further chemical characterization and the remainder to the radioactivity flow detector for direct quantitation. A direct comparison of the chromatograms and the radioactivity counting efficiencies of these three techniques is presented.Entities:
Mesh:
Substances:
Year: 1982 PMID: 6816808 DOI: 10.1093/chromsci/20.11.523
Source DB: PubMed Journal: J Chromatogr Sci ISSN: 0021-9665 Impact factor: 1.618