Literature DB >> 6812378

An oxygen-consuming model lung for evaluation of anaesthetic circuits.

O Stenqvist, H Sonander, B Löfström, K Nilsson.   

Abstract

We have constructed an oxygen-consuming model lung based on pure combustion of hydrogen. Besides consuming oxygen the model produces water vapour, and carbon dioxide is delivered to the lung, thus mimicking carbon dioxide production. Oxygen consumption and carbon dioxide production can be set at desired values. The model lung can be used for investigations on and evaluation of all types of anaesthesia circuits. Results in a circle system with carbon dioxide absorption show that FIO2 is related to the fresh gas flow. The end tidal CO2 level is correlated to the alveolar ventilation and cannot be used for predicting FIO2.

Entities:  

Mesh:

Substances:

Year:  1982        PMID: 6812378     DOI: 10.1111/j.1399-6576.1982.tb01774.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  2 in total

1.  Validation and clinical application of a first order step response equation for nitrogen clearance during FRC measurement.

Authors:  Gary Choncholas; Soren Sondergaard; Erkki Heinonen
Journal:  J Clin Monit Comput       Date:  2007-11-15       Impact factor: 2.502

2.  Improved response time with a new miniaturised main-stream multigas monitor.

Authors:  Mattias Berggren; Nasser Hosseini; Krister Nilsson; Ola Stenqvist
Journal:  J Clin Monit Comput       Date:  2009-10-10       Impact factor: 2.502

  2 in total